QUANTIFICATION OF SCATTERED RADIATION TO ADULT FEMALE BREASTS DURING LUMBOSACRAL X-RAY AND HEAD COMPUTED TOMOGRAPHY IN A NEGROID POPULATION

Mohammed Sidi, Chukwuemeka H Elugwu

Abstract


Abstract

Background: Deterministic and stochastic effects deleterious side effects arising from the use of ionizing radiation in diagnostic radiology, radiation therapy, and nuclear medicine.  The breast is a radiosensitive organ that is often ‘shielded’ with a simple cotton gown when contiguous or fairly distant anatomical regions are imaged radiographically.

 

Objective: To quantify the amount of scattered radiation to breasts during x-ray and CT investigations with a view to finding evidence for apparel shielding during such procedures in the locality.

 

Methods 

Thermoluminescent dosimeters were used to quantify scattered radiation dose in fifty patients each for computed tomography (CT) and computed radiography (CR). Absorbed dose (mGy) were subjected to statistical analysis. Central tendencies and dispersion were noted. Furthermore, a paired-sample t-test was done to test for statistically significant difference in mean absorbed dose by both breasts. Difference found justified the necessity for shielding during radiographic examinations.

 

Results

Scattered radiation reached the breasts in both procedures, and with a range of 1.28 – 12.6 mGy (CT) and 1.02 – 3.63 mGy. There was enormous reduction in dose between unshielded and shielded breasts in CT (113 %) and CR (32.2 %), with the difference being statistically significant (p < 0.05),

 

 

Conclusion

Scattered radiation to the breasts were greatly and significantly reduced when shielding was applied. We strongly urge radiographers to practice shielding of contiguous and distant organs during procedures involving ionizing radiation.

 

Keywords

Scattered radiation, computed tomography, computed radiography, head, lumbosacral spine, x-ray


Full Text:

PDF

References


Elshami W, Abuzaid MM, and Tekin HO.. Effectiveness of Breast and Eye Shielding During Cervical Spine Radiography: An Experimental Study. Risk Manag Healthc Policy. 2020; 13: 697–704.

Joseph DZ, Nzotta CC, Skam JD, Umar MS, Musa DY. Diagnostic reference levels for mammography examinations in North Eastern Nigeria. Diagnostic Reference Levels for Mammography Examinations in North Eastern Nigeria. Afr J Med Health Sci 2018; 17:54-9

Adejoh T, Elugwu CH, Ezugwu EE, Sidi M, Aronu ME. A single-centre experience of radiation dose mitigation in computed radiography using dose charts. Pakistan Journal of Radiology, 2020; 30(1): 25 – 33

Revel M-P, Fitton I, Audureau E, Benzakoun J, Lederlin M, Chabi M-L, Rousset P. Breast Dose Reduction Options During Thoracic CT: Influence of Breast Thickness. American Journal of Roentgenology, 2015; 204:W421-W428

Karami, V., Zabihzadeh, M., and Keshtkar, M., (2017) Evaluation of the entrance surface dose and radiation dose to radiosensitive organs in peadiatric pelvic radiography. Int. J of pead; 5, 5013-5022.

Smith-Bindman R, Lipson J, Marcus Ralph, Kim KP, Mahesh M, Gould R, Berrinton de Gonzalez A, Miglioretti DL. Radiation dose associated with common computed tomography examinations and the associated lifetime attributable risk of cancer. Arch intern Med., 2009; 14(169): 2078 – 2086 doi:10.1001/archinternmed.2009.427

Ekpo, E; Adejoh, T; Akwo, J; Onwujekwe, EC; Ali, AA; Abba, M; Adesina, K; Omiyi, D; Chiegwu, HC. Diagnostic Reference Levels for Common Computed Tomography (CT) Examinations: Results from the first Nigerian Nationwide Dose Survey. Journal of Radiological Protection, 2018; 38:525 – 535

Parker MS, Kelleher NM, Hoots JA, Chung JK, Fatouros PP, Benedict SH. Absorbed radiation dose of the female breast during diagnostic multidetector chest CT and dose reduction with a tungsten-antimony composite breast: preliminary results. Clin Radiol., 2008; 63(3):278 – 288

Gunn ML, Kanal KM, Kolokythas O, Anzal Y. Radiation dose to the thyroid gland and breast from mutidetector computed tomography of the cervical spine: does bismuth shielding with and without a cervical collar reduce dose? J Comput Assist Tomogr., 2009; 33(6):987 – 990

Adejoh, T.; Onwujekwe, E.C.; Abba, M.; Ali, A.M; Imo, A.S.; Nzotta, C.C.; Chiegwu, H.U. Computed Tomography Scanner Census and Adult Head Dose in Nigeria. The Egyptian Journal of Radiology and Nuclear Medicine, 2018; 49: 66–70

Collenti MC, Micheli OA, Lee KH. To Shield or Not to Shield: Application of Bismuth Breast Shields. American Journal of Roentgenology, 2013; 200: 503 – 507 https://www.ajronline.org/doi/full/10.2214/AJR.12.9997

McCollough CH, Wang J, Gould RG, Orton CG. The use of bismuth breast shields for CT should be discouraged. Med. Phys., 2012; 39 (5), 2321 – 2324 [DOI: 10.1118/1.3681014

Hopper KD, King SH, Lobell ME, TenHave TR, and Weaver JS, “The breast: In-plane x-ray protection during diagnostic thoracic CT - shielding with bismuth radioprotective garments,” Radiology, 1997; 205, 853–858

Brnic Z, Vekic B, Hebrang A, Anic P. Efficacy of breast shielding during CT of the head. Eur Radiol., 2003; 13(11):2436 – 2440

Fazel R, Krumholz HM, Wang Y, et al. Exposure to low-dose ionizing radiation from medical imaging procedures. N Engl J Med. 2009;361(9):849–857. doi:10.1056/NEJMoa0901249

BrniÄ Z, VekiÄ B, Hebrang A, AniÄ P. Efficacy of breast shielding during CT of the head. Eur Radiol. 2003;13(11):2436–2440. doi:10.1007/s00330-003-1945-1