EVALUATION OF REJECTED IMAGES IN COMPUTERIZED RADIOGRAPHY (CR)

Dimas Skam Joseph

Abstract


Introduction

Reject image analysis is a quality control check periodically carried out to ensure that optimally diagnostic images are produced. The overall aim ensures that patient radiation dose due to repeat exposure is drastically reduced. In order to ensure continuous optimal quality assurance of the imaging services provided by radiographers to patients, quality control checks such as reject image analysis are put in place to monitor the quality assurance program. The reject image analysis is not a new concept as it has continuously been used by radiographers to check reasons for repeat exposures and the anatomical part of the body mostly affected.

Method

An audit of retrospectively acquired radiographic images of a total of 1740 patients collected from January to March 2020 was performed by two senior radiographers. A data collection chart was used to collect the data according to the causes of repeat exposures with respect to the anatomical body part examined and then the reject image rate was calculated.

Result

The reject image rate (RIR) for January, February and March 2020 are 9.98%, 11.57% and 9.04% with image cut as the major reason for the reject having 3.7%, 3.35% and 4.06% RIR in the respective months. The overall RIR from January to March 2020 is 9.77%.

Conclusion

Though the overall RIR is within recommended limits, continuous reject image analysis will aid in ensuring optimally diagnostic images are continuously produced with minimal repeat exposures and regular continuous professional development (CPD) is recommended for the Radiographers and Radiographer Assistants in order for them not to become rusty with time. The implications for practice is that low RIR is an indicator that optimal radiography services are rendered and also that patients are not exposed to unnecessary radiation. The choice of CPD should be tailored towards the periodic outcome of the reject image analysis especially for young radiographers. As such, a reject image analysis wallet for each radiographer is encouraged in order to identify individual radiographers’ deficiencies.

Keywords: Reject image rate (RIR), repeat exposure, quality control checks, radiation dose,  radiographers


Full Text:

PDF

References


Centers for Disease Control and Prevention (CDC) (last reviewed: December 7, 2015). https://www.cdc.gov/nceh/radiation/alara.html (Acessed March 23rd, 2020)

Joseph D. S., Ibeanu I.G.E, Zakari IY., Joseph D Z., Mustapha N, Olakunle M. Optimization of radiation protection through quality Assurance in some radio-diagnostic facilities in Katsina State, Nigeria. PJR, 2017, 27(4)

Musolino SV, DeFranco J, Schlueck R., 2008, The ALARA principle in the context of a radiological or nuclear emergency, Health Phys, 2008, 94(2), 109–111. doi:10.1097/01.HP.0000285801.87304.3f

Alyousef KA, Alkahtani S, Alessa R and Alruweili H., 2019, Radiograph Reject Analysis in a Large Tertiary Care Hospital in Riyadh, Saudi Arabia, Global Journal on Quality and Safety in Healthcare, 2 (2), 30–33. https://doi.org/10.4103/JQSH.JQSH_24_18

Monfared A. S., Abdi R., Saber M. A. 2007, Repeat analysis program in radiology departments in Mazandaran province- Iran; Impact on population radiation dose, Iran. J. Radiat. Res, 5 (1), 37-40

Atkinson S, Neep M, Starkey D., 2020, Reject rate analysis in digital radiography: an Australian emergency imaging department case study. J Med Radiat Sci, 67, 72– 9. https://doi.org/10.1002/jmrs.343

Owusu-Banahene J., Darko E. O., Hasford F. Addison E. K., OkyereAsirifi J., 2014, Film reject analysis and image quality in diagnostic Radiology Department of a Teaching hospital in Ghana. Journal of Radiation Research and Applied Sciences, 7(4): 589 – 594. https://doi.org/10.1016/j.jrras.2014.09.012

Jones A. K., Heintz P., Geiser W., Goldman L., Jerjian K., Martin M., Peck D., Pfeiffer D., Ranger N., Yorkston J., 2015, Ongoing quality control in digital radiography: Report of AAPM Imaging Physics Committee Task Group 151. International Journal of Medical Physics Research and Practice, 42(11), 6658 - 6670. https://doi.org/10.1118/1.4932623

Essien I. E., Inyang S. O. and Egbe N. O., 2015, Radiographic film reject rate in some hospitals in Akwa Ibom state, Nigeria. JOBARI, 9(4): 206 -211. ISSN: 2395-3438 (P), ISSN: 2395-3446 (O)

https://soundvet.com/5-advantages-of-digital-radiography-every-dvm-needs-to-know/ (Last updated September 8, 2018). (Accessed March 3, 2020)

Arbese M. Y., Abebe T. D., and Mesele B. A., 2018, Determination and analysis of film reject rate at eight selected governmental diagnostic x-ray facilities in Tigray region, Northern Ethiopian. J Med Phys, 43(4): 270–276. doi: 10.4103/jmp.JMP_40_18

Osahon O.D., Asemota O. I., Olajide B. O. and Igbinovia O. J., 2016, Analysis of rejected films of selected examination in a tertiary radiodiagnostic center in Benin City, Edo state, Nigeria. Palgo J.Med.Medical Sci, 3 (1): 38-41

Lin C., Chan P., Huang K., Lu C., Chen Y. and Chen Y. L., 2016, Guidelines for reducing image retakes of general digital radiography. Advances in Mechanical Engineering, 1-6. DOI:10.1177/1687814016644127

Awad F, Al Naem F, Gemea A, Wedaa N, Mohammed Z. and Elser S.,T2021. X-Ray Film Reject Analysis in Radiology Departments of Port Sudan Hospitals. Int J Radiol Imaging Technol, 2021. 7:072. Doi.org/10.23937/2572-3235.1510072.

Adeko OO, Ariba AJ, Olatunji AA, Toyobo OO. Routine chest radiograph in pre-employment medical examination for healthcare workers: Time for a review of the protocol. Niger Postgrad Med J. 2017;24:93-6

Rastegar S, Beigi J, Saeidi E, et al. Reject analysis in digital radiography: A local study on radiographers and students' attitude in Iran 2019. Medical Journal of the Islamic Republic of Iran, 33, 49. DOI: 10.34171/mjiri.33.49.

Andersen E. R., Jorde J., Taoussi N., Yaqoob S. H., Konst B., Seierstad T., V Reject analysis in direct digital radiography. 53 (2), 174-178 https://doi.org/10.1258/ar.2011.110350

Benza C., Damasi-Kasi C., Daniels E. R., Amkongo M. and Nabasenja C., 2018, The causes of reject images in radiology department at a state hospital in Windhoek, Namibia. The South African Radiographer, 56 (1). 35-40

Joseph D. Z, Mohammed S, Samuel S, Abubakar M, Goni M, Itopa R., 2015 Film reject analysis in radiology department of a teaching hospital in north-eastern Nigeria. Nigerian Journal of Medical Imaging and Radiation Therap, 4 (1), 21- 27.

Zewdu M, Kadir E, Berhane M., 2017, Analysis and Economic Implication of X-Ray Film Reject in Diagnostic Radiology Department of Jimma University Specialized Hospital, Southwest Ethiopia, Ethiop J Health Sci, 2007, 27(4):421–426. doi:10.4314/ejhs.v27i4.13

Monfared A. S., Abdi R., Saber M. A., 2007, Repeat analysis program in radiology departments in Mazandaran province- Iran; Impact on population radiation dose. Iran. J. Radiat. Res, 5 (1), 37-40


Refbacks

  • There are currently no refbacks.


© Copyright PJR 2008-