CT Radiation Dose Reduction in Pediatric Cardiovascular Anomalies – Clinical Audit in a Single Centre

Shahjehan Alam, Ummara Siddique Umer, Syed Ghulam Ghaus, Seema Gul, Aman Nawaz Khan, Sadia Gul, Rafiq Ahmed, Shaista Nawaz

Abstract


Purpose:
To reduce radiation dose from Multidetector computed tomography (MDCT) using techniques with low tube voltage and modified image parameters without significant degradation of image quality.

AUDIT TARGET:

Radiation dose reduction   using   80   kV   and   100   kV   protocols   for   MDCT.

 

Materials and Methods:
This is  a prospective analysis with single blind selection of cardiac anomaly patients referred to radiology department of Rehman medical institute Peshawar for cardiac CT scan during Jan 2013 to Dec 2015. Total of 100 patients with age range of 1 week to 16 years were selected with echocardiographic suspicion of cardiovascular anomalies. It was probability sampling. Scan was performed on 128 multislice Toshiba scanner. ECG gated retrospective and prospective scanning was performed using modified tube voltage (80kVp, 100kVp and 120 kVp) and with manual adjustment of low tube current. Radiation dose measurement was done by multiplying conversion factor with dose length product (DLP), which was provided by the CT scanner. Data wa processed using Microsoft Excel 2010. Images were reviewed on 5.1 vitrea workstation using multiplanar and 3D reconstruction. Two radiologists independently assessed subjective quality of the CT images to assess cardiac anomalies and normal anatomical structures.

Results:

Prospective ECG gating significantly reduced the radiation dose 22 mSv (retrospective) to 13.96 mSv (prospective) with standard inbuilt 120kVp setting. Reducing the kVp to 100 further reduced dose to 10.3 mSv and dropped down to 5.46mSv with 80kVp and no significant image distortion. Manually adjusting and reducing the mAs with added filtration reduced the radiation dose to 3.64mSv.

Conclusion:
We conclude from our results that prospective ECG gated cardiac CT, low kVp and mAs values show great potential for substantially reducing radiation dose of cardiac CT angiography. By using different radiation lowering dose techniques i.e. prospective ECG gating technique, low kVp and mAs, we were able to reduce radiation dose by 83.45% (22 to 3.46 mSv)


Full Text:

PDF

References


Yu L, Liu X, Leng S, Kofler JM, Ramirez-Giraldo JC, Qu M, Christner J, Fletcher JG, and McCollough CH. Radiation dose reduction in computed tomography: techniques and future perspective. Imaging Med. 2009 Oct; 1(1): 65–84.

McCollough CH, Primak AN, Braun N, Kofler J, Yu L, and Christner J, PhD. Strategies for Reducing Radiation Dose in CT. Radiol Clin North Am. 2009 Jan; 47(1): 27–40.

Bethesda. National Council on Radiation Protection and Measurements. Ionizing radiation exposure of the population of the United States: National Council on Radiation Protection and Measurements; 1987. NCRP Report No. 93.

Brenner DJ, Hall EJ. Computed tomography-an increasing source of radiation exposure N Engl J Med. 2007 Nov 29; 357(22):2277-84.

Radiation Dose Reduction without Degradation of Low-Contrast Detectability at Abdominal Multisection CT with a Low–Tube Voltage Technique: Phantom Study. Funama Y, Awai K, Nakayama Y, Kakei K, Nagasue N, Shimamura M, Sato N, Sultana S, Morishita S and Yamashita Y. Radiology 2005; 237(3):905-10.

Hausleiter J, Meyer T, Hermann F, Hadamitzky M, Krebs M, Gerber TC, McCollough C, Martinoff S, Kastrati A, Schömig A, Achenbach S.Estimated radiation dose associated with cardiac CT angiography. JAMA. 2009;301(5):500-7.

Committee to Assess Health Risks From Exposure to Low Levels of Ionizing Radiation; Nuclear and Radiation Studies Board; National Research Council of the National Academies. Health Risks From Exposure to Low Levels of Ionizing Radiation, BEIR VII-Phase 2. Washington, DC: National Academies Press; 2006.

Raff GL, Chinnaiyan KM,Share DA, Goraya TY, Kazerooni EA, Moscucci M, Gentry RE, Abidov A. Radiation Dose From Cardiac Computed Tomography Before and After Implementation of Radiation Dose–Reduction Techniques for the Advanced Cardiovascular Imaging Consortium Co-Investigators . JAMA. 2009;301(22):2340-8.

The Royal College of Radiologists. Making the Best Use of Clinical Radiology Services: Referral Guidelines. 6th Edition. London: 2007

Biologic Effects of Ionizing Radiation (BEIR) Report VII: Health risks from exposure to low levels of ionizing radiation. National Academies. 2006.

Brenner D, Elliston C, Hall E, Berdon W. Estimated risks of radiation-induced fatal cancer from pediatric CT. AJR Am J Roentgenol. 2001 Feb; 176(2):289-96.

Arch ME, Frush DP. Pediatric body MDCT: a 5-year follow-up survey of scanning parameters used by pediatric radiologists. AJR Am J Roentgenol. 2008 Aug; 191(2):611-7.

Reston, VA. Practice Guidelines and Technical Standards. ACR practice guideline for diagnostic reference levels in medical x-ray imaging: American College of Radiology. 2008:799–804.

International Electrotechnical Commission. Medical electrical equipment. Part 2–44: Particular requirements for the safety of x-ray equipment for computed tomography. IEC publication No. 60601–60602–60644. Ed. 2.1. Geneva, Switzerland: International Electrotechnical Commission (IEC); 2002.

Valentin J. Managing patient dose in multi-detector computed tomography (MDCT). Ann ICRP. 2007; 37(1):1-79.

Jacobi W. The concept of the effective dose--a proposal for the combination of organ doses. Radiat Environ Biophys 1975 Jun 18; 12(2):101-9.

McCollough CH, Schueler BA. Calculation of effective dose. Med Phys. 2000 May; 27(5):828-37.

International Commission on Radiological Protection: 1990 Recommendations of the International Commission on Radiological Protection (Report 60) Ann. ICRP. 1991;21:1–3.

Siegel MJ, Schmidt B, Bradley D, Suess C, Hildebolt C. Radiation dose and image quality in pediatric CT: effect of technical factors and phantom size and shape. Radiology. 2004 Nov; 233(2):515-22.

McCollough CH., Primak AN, Braun N, Kofler J, Yu L & Christner J. Strategies for Reducing Radiation Dose in CT.Radiologic clinics of North America 2009;47(1): 27–40.

Hausleiter J, Meyer T, Hermann F, Hadamitzky M, Krebs M, Gerber TC, McCollough C, Martinoff S, Kastrati A, Schömig A, Achenbach S. Estimated radiation dose associated with cardiac CT angiography. JAMA. 2009 Feb 4; 301(5):500-7.

Hsieh J, Londt J, Vass M, Li J, Tang X, Okerlund D. Step-and-shoot data acquisition and reconstruction for cardiac x-ray computed tomography. Med Phys. 2006 Nov; 33(11):4236-48.

Stolzmann P, Leschka S, Scheffel H, Krauss T, Desbiolles L, Plass A, Genoni M, Flohr TG, Wildermuth S, Marincek B, Alkadhi H. Dual-source CT in step-and-shoot mode: noninvasive coronary angiography with low radiation dose. Radiology. 2008 Oct; 249(1):71-80.

Scheffel H, Alkadhi H, Leschka S, Plass A, Desbiolles L, Guber I, Krauss T, Gruenenfelder J, Genoni M, Luescher TF, Marincek B, Stolzmann P. Low-dose CT coronary angiography in the step-and-shoot mode: diagnostic performance. Heart. 2008 Sep; 94(9):1132-7.

Kak AC, Slaney M. Principles of Computed Tomographic Imaging. PA, USA: SIAM; 1987.

Bai M, Chen J, Raupach R, Suess C, Tao Y, Peng M. Effect of nonlinear three-dimensional optimized reconstruction algorithm filter on image quality and radiation dose: validation on phantoms. Med Phys. 2009 Jan; 36(1):95-7.

Funama Y, Awai K, Nakayama Y, Kakei K, Nagasue N, Shimamura M, Sato N, Sultana S, Morishita S, Yamashita Y. Radiation dose reduction without degradation of low-contrast detectability at abdominal multisection CT with a low-tube voltage technique: phantom study. Radiology. 2005 Dec; 237(3):905-10.

Nakayama Y, Awai K, Funama Y, Hatemura M, Imuta M, Nakaura T, Ryu D, Morishita S, Sultana S, Sato N, Yamashita Y. Abdominal CT with low tube voltage: preliminary observations about radiation dose, contrast enhancement, image quality, and noise. Radiology. 2005 Dec; 237(3):945-51.

Schindera ST, Nelson RC, Mukundan S Jr, Paulson EK, Jaffe TA, Miller CM, DeLong DM, Kawaji K, Yoshizumi TT, Samei E. Hypervascular liver tumors: low tube voltage, high tube current multi-detector row CT for enhanced detection--phantom study. Radiology. 2008 Jan; 246(1):125-32.

Fletcher JG, Takahashi N, Hartman R, Guimaraes L, Huprich JE, Hough DM, Yu L, McCollough CH. Dual-energy and dual-source CT: is there a role in the abdomen and pelvis? Radiol Clin North Am. 2009 Jan; 47(1):41-57.

Gibbons RJ, Araoz PA, Williamson EE. The year in cardiac imaging. J Am Coll Cardiol. 2009 Jan 6; 53(1):54-70.

Siegel MJ, Schmidt B, Bradley D, Suess C, Hildebolt C. Radiation dose and image quality in pediatric CT: effect of technical factors and phantom size and shape. Radiology. 2004 Nov; 233(2):515-22.