DIAGNOSTIC REFERENCE LEVELS (DRLs) For COMPUTED TOMOGRAPHY (CT) EXAMINATIONS IN NORTH EASTERN NIGERIA

Joseph Dlama Zira, Christian Chukwuemeka Nzotta, Joseph Dimas Skam

Abstract


Background: Diagnostic reference levels (DRLs) for computed tomography (CT) examination is important for dose optimization. The are suggested action levels and reference guide for achieving radiation protection among patients.

Objective of the study: To establish DRL for CT examination in north eastern Nigeria and to compare it with other established work.

Methodology: Prospective cross-sectional study conducted in two hospitals in north eastern Nigeria. One hundred and eighty subjects were recruited for the study. Computed tomography dose index (CTDI vol), dose length product (DLP) and Scan parameters were obtained. Weight, height and body mass index (BMI) were recorded. Student T-test was used to compare the relationship between the mean CTDIvol obtained in the two centers, Pearson’s correlation was used to determine the relationship between the dosimetric and anthropotechnical parameters. Statistical significance was set at P<0.05.

Results: The DRL values obtained in this work were 67.90 mGy, 18.83 mGy and 19.20 mGy for head CT, Chest CT and CT abdomen respectively. There was no statistical significant relationship (p>0.05) between CTDIvol with thickness, weight, height and BMI for head CT and abdominal CT. However, CTDI for Chest CT show statistical significant relationship (p<0.05) with weight and height.

Conclusion: DRL was higher when compared with international values. There is need for optimization of radiology practice in North Eastern Nigeria.

Key words: X-rays, computed tomography, diagnostic reference levels, CTDIvol, dose length product, dose, optimization.


Full Text:

PDF

References


Reference

Chougule A. Reference dose in radiological imaging. Polish Journal of Medical Physics and Engeneering 2005;11:115-126.

Hall EJ, Brenner DJ. Cancer risk from diagnostic radiology. British Journal of Radiology. 2008;81 362-378: doi.org/10.1259/bjr/01948454. PubMed PMID: 18440940.

Mettler Jr FA, Huda W, Yoshizumi TT,Mahesh M, Effective doses in radiology and diagnostic nuclear medicine: A Catalog 1. Radiology. 2008; 248:254-263. Doi.org/10.1148/radiol.2481071451.

Saravanakumar A, Vaideki K, Govindarajan K, Davanand B, Jayakumar S and Sharma SD. Establishment of computed tomography reference levels in selected procedures in south India. International Journal of Radiation Research,2016; 14(4): 341-347

Donald L, Deukwoo K, Grant H. Reference Levels for Patients Radiation Doses in Interventional Radiology: Proposed Initial Values for U.S. Practice.Radiology,2010; 253 (3): 753-764.

Jeska S, Goeffrey K, Mark .A, Ian K, Jedidah M. Patients Radiation exposure during general fluoroscopy examinations. Journal of Applied Clinical Medical Physics ,2014; 15(2): 1-10.

Wallace A B. The Implementation of Diagnostic Reference Levels to Australian Radiology Practice. Journal of Medical. Imaging and Radiation Oncology, 2010; 54 (5): 465- 471

Carroll. E and Brennan P. C. Radiation doses for barium enema and barium meal examinations in Ireland: Potential diagnostic reference levels. British journal of Radiology, 2014; (10): 1259-1264

ICRP (2000)“Recommendation of International Commission on Radiological Protection” ICRP Publication64 Ann ICRP 1991 21 (1-3) Pergamum Press, Oxford UK.

ICRP, (2011). Diagnostic reference levels in medical imaging: Review and additional advice. Annals of ICRP,31(4), pp. 33-35.

Hart D, M.C. Hillier, B.F. Wall, P.C. Shrimpton and D.Bungay.Doses to Patient from Medical X-ray Examinations in the UK - 1995 Review. National Radiological Protection Board Publication NRPB-R 289.2011.

Muhammed K, Cyril S and Francis H. Determination of computed tomography Diagnostic reference levels in North Central Nigeria. Pacific Journal of Science and Technology. 2016; 17(2): 341-349.

Joseph D, Obetta C, Nkubli F, Geofrey L, Laushugno S, Yabwa D. Rationale for implementing dose reference levels as a quality assurance tool in medical radiography in Nigeria. IOSR Journal of Dental and Medical Sciences. 2014; 13(12): 41-45

Jessen KA, Shrimpton PC, Geleijns J, Panzer W, Tosi G. Dosimetry for optimization of patient protection in computed tomography. Applied radiation isotopes 1999; 50: 165-72. Doi.org/10.1016/S0969-8043(98)00024-4. PubMed PMID:10028635.

McCollough C,Cody D, Edyvean S, Geise R, Gould B, Keat N, et al. The measurement, reporting and management of radiation dose in CT. Report of AAPM Task group. 2008;23:1-28.

Abdullahi M, Shittu H, Arabisala A, Eshiett P, Joseph D.Z, Richard I, Kpaku G. (2015). Diagnostic Reference Level for Adult Brain Computed Tomography Scans: A Case Study of a Tertiary Health Care Center in Nigeria. Journal of Dental and Medical Sciences (IOSR-JDMS),14, 1 (2): 66-75.

Joseph Z and Nzotta CC.. The need to establish dose reference levels for radiological examinations in Nigeria: Radiographers role. Nigerian Journal of Medical Imaging and Radiation Therapy, 2016;5(1) : 25-39.

Godwin I, Racheal O. Radiation Doses in computed Tomography: Need for Optimization and application of dose reference levels in Nigeria. West African Journal of Radiology,2014;21(1): 1-6