PERTINENT QC TESTS FOR GAMMA CAMERA PERFORMANCE IN A BUSY NUCLEAR MEDICINE DEPARTMENT: AN AUDIT

Authors

  • Muhammad Yaseen Department of Radiology, Lady Reading Hospital, Pakistan
  • Amir Ali Department of Radiology, Lady Reading Hospital, Pakistan
  • Tahira Nishtar Department of Radiology, Lady Reading Hospital, Pakistan
  • Badar Jahan Department of Surgery, SMBBMC, Lyari, Karachi, Pakistan

Abstract

Background: In nuclear medicine imaging comprehensive quality controls procedures are used to ensure the accuracy and reproducibility gamma camera. However, these are time consuming. In practice, however, less time-consuming and less rigorous procedures often suffice for day-to-day QC. We are presenting an audit of QC of dual head gamma camera of aNuclear Medicine Department of a tertiary care hospital in Karachi.

Material and Method: This audit was performed at Nuclear Medicine Department of Liaquat National Hospital (LNH) Karachi, Pakistan from 1st October 2016 till 31 Dec 2015. We perform background count, energy peaking, % energy resolution (full width half maximum; FWHM), % central and uniform field of view (% CFOV and % UFOV), total counts and count rate.

Results: Mean background counts (Kcnts) detected by Detector-1 was 0.24 and for Detector-2 was 0.22 which were well within NEMA benchmark but values of detectors were significantly different.  Mean % energy resolution (%FWHM) for Detector-1 was 09.89 and for Detector-2 was 10.364 (within NEMA standards) but significantly difference between 02 detectors. Mean count rate (Kcts/s) detected by Detector-1 was 14.27 and for Detector-2 was 15.91 (well within NEMA benchmark) but significant difference between counts rate of 02 detectors was found.

Conclusion: We conclude that less time consuming and convenient procedures like background count, energy resolution and mean count rate measurement are sensitive tests for daily QC of a busy nuclear medicine department.

 

Key Words: Gamma camera; Quality control; uniformity; energy resolution; Electronic drift

 

 

 

References

. Busemann, E. et al.: Routine quality control recommendations for nuclear medicine instrumentation. 2010; 37:662-71.

. Wegst AV, Erickson JJ. Quality control computer interface /scintillation cameras. Med Phys World 1991; 7:13-30

. Zanzonico P. Routine Quality Control of Clinical Nuclear Medicine Instrumentation: A Brief Review. J Nucl Med. 2008; 49: 1114–31.

. Nichols KJ, Bacharach SL, Bergmann SR, et al. Instrumentation quality assurance and performance. J Nucl Cardiol 2006;13:e25–e41

Downloads

Published

02-04-2017

How to Cite

Yaseen, M., Ali, A., Nishtar, T., & Jahan, B. (2017). PERTINENT QC TESTS FOR GAMMA CAMERA PERFORMANCE IN A BUSY NUCLEAR MEDICINE DEPARTMENT: AN AUDIT. Pakistan Journal of Radiology, 27(2). Retrieved from https://www.pakjr.com/index.php/PJR/article/view/620

Issue

Section

Original Articles